To assess the feasibility and accuracy in measuring left ventricular (LV) end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (EF) with Siemens single-beat real-time 3D transthoracic echocardiography. The LV volumes and EF were measured in 3D datasets acquired by six imaging modes (time-1-harmonic (T1H), time-1-fundamental, time-2-harmonic, time-2-fundamental, space-1-harmonic (S1H), and space-1-fundamental) in 41 patients using the automated contouring algorithm and compared with manually corrected 3DE QLAB measurements. The main determinates of the temporal and spatial resolutions of 3D datasets acquired were the fundamental and harmonic modes. Consequently, the S1H mode had the lowest volume rate and highest spatial resolution. Compared with the 3DE QLAB analysis, the S1H mode resulted in the best LV volumes and EF estimates in all patients (0 ± 10 % for EF, -7 ± 44 ml for EDV, -7 ± 39 ml for ESV) and in the 10 patients with correct LV contour tracking according to a visual assessment from the multiplanar reconstruction views in all six modes (0 ± 9 % for EF, -3 ± 23 ml for EDV, -2 ± 14 ml for ESV). The T1H mode was the best alternative. Overall 28 patients (68 %) could be analysed automatically and satisfyingly with the S1H and T1H modes: 0 ± 8 % (EF), 0 ± 27 ml (EDV) and -1 ± 16 ml (ESV). The accuracy of the Siemens automated RT-3D algorithm in measuring LV volumes and EF is significantly influenced by the different imaging modes. The S1H mode may be the preferred 3D acquisition mode, supplemented by the T1H mode in enlarged LVs that do not fit in the S1H acquisition sector.

, ,
doi.org/10.1007/s10554-013-0327-2, hdl.handle.net/1765/62848
International Journal of Cardiovascular Imaging
Department of Cardiology

Ren, B., Vletter, W., Vletter-McGhie, J., Soliman, O. I. I., & Geleijnse, M. (2013). Single-beat real-time three-dimensional echocardiographic automated contour detection for quantification of left ventricular volumes and systolic function. International Journal of Cardiovascular Imaging, 1–8. doi:10.1007/s10554-013-0327-2