Exhaled nitric oxide measurements with dynamic flow restriction in children aged 4-8 yrs


Article
Related Files
asset icon
(12412684.pdf, 0.2MB)

Fractional exhaled nitric oxide concentration (FENO) depends on exhalation flow; however, children often are unable to perform controlled flow procedures. Therefore, a device was developed for off-line FENO sampling, with dynamic flow restriction (DFR). The authors compared off-line with on-line FENO, assessed feasibility, and obtained normal values for FENO in children aged 4-8 yrs. Subjects inhaled nitric oxide (NO)-free air and exhaled into the device, where DFR kept exhalation flow constant at 50 mL x s(-1). Dead space air was discarded. Exhaled air was collected in a 150 mL mylar balloon. On-line measurements were performed and values compared with off-line FENO in 19 adult volunteers. Seventy-nine children performed off-line sampling. All samples were analysed with a chemiluminescence NO-analyser. Normal values were obtained in 34 healthy children. There was an excellent correlation between on- and off-line values. Bland and Altman plots showed good agreement between on- and off-line FENO. Seventy-four out of 79 children were able to perform a correct off-line procedure. Geometric mean+/-SEM FENO in healthy children was 4.9+/-1.2 parts per billion (ppb) for male children and 7.6+/-1.1 ppb for female children. It can be concluded that off-line fraction of exhaled nitric oxide measurements with dynamic flow restriction are feasible in young children and correspond to on-line values.



Keywords


Automatically Extracted Terms
  • children
  • nitric
  • oxide
  • nitric oxide
  • value
  • measurement
  • correlation
  • flow restriction
  • nitric oxide measurements
  • study
  • author
  • respir
  • feno values
  • exhalation
  • ambient
  • agreement
  • symptom
  • sampling
  • method
  • space