Histogram-based selective deblurring to improve computed tomography imaging of calcifications
January 2007
Article
| Related Files |
|---|
|
Redirect to publisher's version
(publisher's version.url.txt, 52 bytes) |
OBJECTIVES: Computed tomography (CT) imaging of small high-density structures, eg, calcifications, is hampered by image blur. This study aims to deconvolve calcifications in the transverse and longitudinal directions while avoiding noise amplification and edge-ringing artifacts in the surrounding low-density structures. MATERIALS AND METHODS: A method referred to as histogram-based selective deblurring (HiSD) has been developed to generate a restored image by combining the low-intensity (ie, Hounsfield Units) information of the original image with the high-intensity information of the deconvolved image. HiSD is evaluated on phantom and in vitro atherosclerotic plaque CT images by comparing the original and restored images with their corresponding reference micro-CT images both qualitatively and quantitatively. RESULTS: HiSD reduces calcification blur in the transverse and longitudinal directions without introducing noise and ringing-artifacts in the surrounding tissues. Calcification area and volume measurements are significantly improved in the restored images (reducing on average overestimation by 32% and 83%, respectively). CONCLUSIONS: HiSD significantly improves CT visualization and quantification of small high-density structures imaged in vitro.
- Humans
- Reproducibility of Results
- Algorithms
- Phantoms, Imaging
- Imaging, Three-Dimensional
- *Tomography, X-Ray Computed
- Calcinosis/*radiography
- Atherosclerosis/*radiography
- Radiographic Image Enhancement/*methods