2017-10-31
The directivity of piezoelectric matrix transducer elements mounted on an ASIC
Publication
Publication
Over the last decade, clinical studies show a strong interest for real-time 3D imaging. This calls for ultrasound probes with high-element-count 2D matrix transducer array, interfaced to an imaging system using an in-probe Application Specific Integrated Circuit (ASIC) that takes care of element selection, signal amplification, sub-array beamforming, etc. Since the ASIC is based on silicon and is mounted directly behind the transducer elements, it can be regarded as a rigid plate that can sustain travelling waves, which effectively lead to acoustical cross-talk between the elements. We hypothesize that the cross-talk can be diminished by reducing the thickness of the ASIC and using a proper backing, and we investigate this by simulation.
Additional Metadata | |
---|---|
doi.org/10.1109/ULTSYM.2017.8092118, hdl.handle.net/1765/103740 | |
2017 IEEE International Ultrasonics Symposium, IUS 2017 | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Shabanimotlagh, M., Raghunathan, S. B., Daeichin, V., Kruizinga, P., Vos, R., Pertijs, M., … Verweij, M. (2017). The directivity of piezoelectric matrix transducer elements mounted on an ASIC. In IEEE International Ultrasonics Symposium, IUS. doi:10.1109/ULTSYM.2017.8092118 |