Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff
Publication
Publication
American Journal of Hypertension , Volume 22 - Issue 4 p. 378- 383
Background: The Finapres methodology offers continuous measurement of blood pressure (BP) in a noninvasive manner. The latest development using this methodology is the Nexfin monitor. The present study evaluated the accuracy of Nexfin noninvasive arterial pressure (NAP) compared with auscultatory BP measurements (Riva-Rocci/Korotkoff, RRK). Methods: In supine subjects NAP was compared to RRK, performed by two observers using an electronic stethoscope with double earpieces. Per subject, three NAP-RRK differences were determined for systolic and diastolic BP, and bias and precision of differences were expressed as median (25th, 75th percentiles). Within-subject precision was defined as the (25th, 75th percentiles) after removing the average individual difference. Results: A total of 312 data sets of NAP and RRK for systolic and diastolic BP from 104 subjects (aged 18-95 years, 54 males) were compared. RRK systolic BP was 129 (115, 150), and diastolic BP was 80 (72, 89), NAP-RRK differences were 5.4 (-1.7, 11.0) mm Hg and -2.5 (-7.6, 2.3) mm Hg for systolic and diastolic BP, respectively; within-subject precisions were (-2.2, 2.3) and (-1.6, 1.5) mm Hg, respectively. Conclusion: Nexfin provides accurate measurement of BP with good within-subject precision when compared to RRK.
Additional Metadata | |
---|---|
, , , , , , , , , , , , , , , , , , | |
doi.org/10.1038/ajh.2008.368, hdl.handle.net/1765/16564 | |
American Journal of Hypertension | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Eeftinck Schattenkerk, D., van Lieshout, J., van den Meiracker, A., Wesseling, K., Blanc, S., Wieling, W., … Westerhof, B. (2009). Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. American Journal of Hypertension, 22(4), 378–383. doi:10.1038/ajh.2008.368 |