Dose reduction in spiral CT coronary angiography with dual-source equipment. Part I. A phantom study applying different prospective tube current modulation algorithms
Publication
Publication
La Radiologia Medica , Volume 114 - Issue 7 p. 1037- 1052
Purpose: The authors sought to compare different algorithms for dose reduction in retrospectively echocardiographically (ECG)-gated dual-source computed tomography (CT) coronary angiography (DSCT-CA) in a phantom model. Materials and methods: Weighted CT dose index (CTDI) was measured by using an anthropomorphic phantom in spiral cardiac mode (retrospective ECG gating) at five pitch values adapted with two heart-rate-adaptive ECG pulsing windows using four algorithms: narrow pulsing window, with tube current reduction to 20% (A) and 4% (B) of peak current outside the pulsing window; wide pulsing window, with tube current reduction to 20% (C) and 4% (D). Each algorithm was applied at different heart rates (45, 60, 75, 90, 120 bpm). Results: Mean CTDI volume (CTDIvol) was 36.9±9.7 mGy, 23.9±5.6 mGy, 49.7±16.2 mGy and 38.5±12.3 mGy for A, B, C and D, respectively. Consistent dose reduction was observed with protocols applying the 4% tube current reduction (B and D). Using the conversion coefficient for the chest, the mean effective dose was the highest for C (9.6 mSv) and the lowest for B (4.6 mSv). Heart-ratedependent pitch values (pitch=0.2, 0.26, 0.34, 0.43, 0.5) and the use of heart-rate-adaptive ECG pulsing windows provided a significant decrease in the CTDIvol with progressively higher heart rates (45, 60, 75, 90, 120 bpm), despite using wider pulsing windows. Conclusions: Radiation exposure with DSCT-CA using a narrow pulsing window significantly decreases when compared with a wider pulsing window. When using a protocol with reduced tube current to 4%, the radiation dose is significantly lower.
Additional Metadata | |
---|---|
, , , | |
doi.org/10.1007/s11547-009-0437-z, hdl.handle.net/1765/24231 | |
La Radiologia Medica | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Martini, C., Palumbo, A., Maffei, E., Rossi, A., Rengo, M., Malagó, R., … Cademartiri, F. (2009). Dose reduction in spiral CT coronary angiography with dual-source equipment. Part I. A phantom study applying different prospective tube current modulation algorithms. La Radiologia Medica, 114(7), 1037–1052. doi:10.1007/s11547-009-0437-z |