Diagnostic accuracy of second-generation dual-source computed tomography coronary angiography with iterative reconstructions: a real-world experience
Publication
Publication
La Radiologia Medica , Volume 117 - Issue 5 p. 725- 738
Purpose: The authors evaluated the diagnostic accuracy of second-generation dual-source (DSCT) computed tomography coronary angiography (CTCA) with iterative reconstructions for detecting obstructive coronary artery disease (CAD). Materials and methods: Between June 2010 and February 2011, we enrolled 160 patients (85 men; mean age 61.2±11.6 years) with suspected CAD. All patients underwent CTCA and conventional coronary angiography (CCA). For the CTCA scan (Definition Flash, Siemens), we use prospective tube current modulation and 70-100 ml of iodinated contrast material (Iomeprol 400 mgI/ ml, Bracco). Data sets were reconstructed with iterative reconstruction algorithm (IRIS, Siemens). CTCA and CCA reports were used to evaluate accuracy using the threshold for significant stenosis at ≥50% and ≥70%, respectively. Results: No patient was excluded from the analysis. Heart rate was 64.3±11.9 bpm and radiation dose was 7.2±2.1 mSv. Disease prevalence was 30% (48/160). Sensitivity, specificity and positive and negative predictive values of CTCA in detecting significant stenosis were 90.1%, 93.3%, 53.2% and 99.1% (per segment), 97.5%, 91.2%, 61.4% and 99.6% (per vessel) and 100%, 83%, 71.6% and 100% (per patient), respectively. Positive and negative likelihood ratios at the per-patient level were 5.89 and 0.0, respectively. Conclusions: CTCA with second-generation DSCT in the real clinical world shows a diagnostic performance comparable with previously reported validation studies. The excellent negative predictive value and likelihood ratio make CTCA a first-line noninvasive method for diagnosing obstructive CAD.
Additional Metadata | |
---|---|
, , , , , , | |
doi.org/10.1007/s11547-011-0754-x, hdl.handle.net/1765/33211 | |
La Radiologia Medica | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Maffei, E., Martini, C., Rossi, A., Mollet, N., Lario, C., Castiglione Morelli, M., … Cademartiri, F. (2012). Diagnostic accuracy of second-generation dual-source computed tomography coronary angiography with iterative reconstructions: a real-world experience. La Radiologia Medica, 117(5), 725–738. doi:10.1007/s11547-011-0754-x |