In protein and peptide mass spectrometry in which profiling of peaks is involved, their masses and intensities are important characteristics. Because of the relative low reproducibility of peak intensities associated with complex samples in matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOFMS), it is difficult to accurately assess the number of peaks and their intensities. In this study we evaluate these two characteristics for tryptic digests of cerebrospinal fluid. We observed that the reproducibility of peak intensities was relatively poor (CV = 42%) and that additional normalization or spiking did not lead to a large improvement (CV = 30%). Moreover, at least seven mass spectra per sample were required to obtain a reliable peak list. An improvement of the sensitivity (i.e., eventually more peaks are detected) is observed if more replicates per sample are measured. We conclude that the reproducibility and sensitivity of peptide profiling can be significantly improved by a combination of measuring at least seven spectra per sample with a dichotomous scoring of the intensities. This approach will aid the analysis of large numbers of mass spectra of patient samples in a reproducible way for the detection and validation of candidate biomarkers. Copyright

doi.org/10.1002/rcm.1864, hdl.handle.net/1765/57571
Rapid Communications in Mass Spectrometry