2006-08-01
Roentgen stereophotogrammetric analysis: An accurate tool to assess stent-graft migration
Publication
Publication
Journal of Endovascular Therapy , Volume 13 - Issue 4 p. 468- 475
Purpose: To evaluate in an in vitro model the feasibility and accuracy of Roentgen stereophotogrammetric analysis (RSA) versus computed tomography (CT) for the ability to detect stent-graft migration. Methods: An aortic model was constructed from a 22-mm-diameter Plexiglas tube with 6-mm polytetrafluoroethylene inlays to mimic the renal arteries. Six tantalum markers were placed in the wall of the aortic tube proximal to the renal arteries. Another 6 markers were added to a Gianturco stent, which was cast in Plexiglas and placed inside the aorta and fixed to a micromanipulator to precisely control displacement of the stent along the longitudinal axis. Sixteen migrations were analyzed with RSA software and compared to the micromanipulator. Thirty-two migrations were measured by 3 observers from CT images acquired with 16×0.5-mm beam collimation and reconstructed with a 0.5-mm slice thickness and a 0.4-mm reconstruction interval. Measurements were made with Vitrea postprocessing software using a standard clinical protocol and central lumen line reconstruction. Results of CT were also compared to the micromanipulator. Results: The mean RSA measurement error compared to the micromanipulator was 0.002±0.044 mm, and the maximum error was 0.10 mm. There was no statistically significant interobserver variability for CT (p=0.17). The pooled mean (maximum) measurement error of CT was 0.14±0.29 (1.00) mm, which was significantly different from the RSA measurement error (p<0.0001). Conclusion: Detection of endograft migration by RSA is feasible and was significantly more accurate than CT in this nonpulsatile in vitro model.
Additional Metadata | |
---|---|
, , , , , , , | |
doi.org/10.1583/06-1842MR.1, hdl.handle.net/1765/62218 | |
Journal of Endovascular Therapy | |
Organisation | Biophysical Genomics, Department Cell Biology & Genetics |
Koning, O.H.J, Oudegeest, O.R, Valstar, E.R, Garling, E.H, van der Linden, E, Hinnen, J.-W, … van Bockel, J.H. (2006). Roentgen stereophotogrammetric analysis: An accurate tool to assess stent-graft migration. Journal of Endovascular Therapy, 13(4), 468–475. doi:10.1583/06-1842MR.1
|