Rapid and accurate measurement of left ventricular function with a new second-harmonic fast-rotating transducer and semi-automated border detection
Publication
Publication
Echocardiography: a journal of cardiovascular ultrasound and allied techniques , Volume 23 - Issue 6 p. 447- 454
Measurement of left ventricular (LV) volume and function are the most common clinical referral questions to the echocardiography laboratory. A fast, practical, and accurate method would offer important advantages to obtain this important information. To validate a new practical method for rapid measurement of LV volume and function. We developed a continuous fast-rotating transducer, with second-harmonic capabilities, for three-dimensional echocardiography (3DE). Fifteen cardiac patients underwent both 3DE and magnetic resonance imaging (reference method) on the same day. 3DE image acquisition was performed during a 10-second breath-hold with a frame rate of 100 frames/sec and a rotational speed of 6 rotations/sec. The individual images were postprocessed with Matlab software using multibeat data fusion. Subsequently, with these images, 12 datasets per cardiac cycle were reconstructed, each comprising seven equidistant cross-sectional images for analysis in the new TomTec 4DLV analysis software, which uses a semi-automated border detection (ABD) algorithm. The ABD requires an average analysis time of 15 minutes per patient. A strong correlation was found between LV end-diastolic volume (r = 0.99; y = 0.95x - 1.14 ml; SEE = 6.5 ml), LV end-systolic volume (r = 0.96; y = 0.89x + 7.91 ml; SEE = 7.0 ml), and LV ejection fraction (r = 0.93; y = 0.69x + 13.36; SEE = 2.4%). Inter- and intraobserver agreement for all measurements was good. The fast-rotating transducer with new ABD software is a dedicated tool for rapid and accurate analysis of LV volume and function.
Additional Metadata | |
---|---|
, , | |
doi.org/10.1111/j.1540-8175.2006.00241.x, hdl.handle.net/1765/65504 | |
Echocardiography: a journal of cardiovascular ultrasound and allied techniques | |
Organisation | Department of Cardiology |
Krenning, E., Voormolen, M., van Geuns, R. J., Vletter, W., Lancée, C., de Jong, N., … Roelandt, J. (2006). Rapid and accurate measurement of left ventricular function with a new second-harmonic fast-rotating transducer and semi-automated border detection. Echocardiography: a journal of cardiovascular ultrasound and allied techniques, 23(6), 447–454. doi:10.1111/j.1540-8175.2006.00241.x |