Rationale We aimed to test the interscan variation of semiautomatic volumetry of subsolid nodules (SSNs), as growth evaluation is important for SSN management. Methods From a lung cancer screening trial all SSNs that were stable over at least 3 months were included (N=44). SSNs were quantified on the baseline CT by two observers using semi-automatic volumetry software for effective diameter, volume, and mass. One observer also measured the SSNs on the second CT 3 months later. Interscan variation was evaluated using Bland-Altman plots. Observer agreement was calculated as intraclass correlation coefficient (ICC). Data are presented as mean (± standard deviation) or median and interquartile range (IQR). A Mann-Whitney U test was used for the analysis of the influence of adjustments on the measurements. Results Semi-automatic measurements were feasible in all 44 SSNs. The interscan limits of agreement ranged from -12.0 % to 9.7 % for diameter, -35.4 % to 28.6 % for volume and - 27.6 % to 30.8 % for mass. Agreement between observers was good with intraclass correlation coefficients of 0.978, 0.957, and 0.968 for diameter, volume, and mass, respectively. Conclusion Our data suggest that when using our software an increase in mass of 30 % can be regarded as significant growth.

, , , ,
doi.org/10.1007/s00330-014-3478-1, hdl.handle.net/1765/85393
European Radiology: journal of the European Congress of Radiology
Department of Pulmonology

Scholten, E., de Jong, P., Jacobs, C., van Ginneken, B., van Riel, S., Willemink, M. J., … Gietema, H. (2015). Interscan variation of semi-automated volumetry of subsolid pulmonary nodules. European Radiology: journal of the European Congress of Radiology, 25(4), 1040–1047. doi:10.1007/s00330-014-3478-1