2015
Inappropriate ICD shocks after collision with a pole
Publication
Publication
Nederlands Tijdschrift voor Geneeskunde , Volume 159 p. A9642
A 60-year-old man experienced inappropriate ICD shocks after collision with a pole. Cine-fluoroscopy demonstrated lead fracture of the proximal lead, which was not so clearly visible on the chest radiograph. He underwent lead extraction and implantation of a new ICD system.
Additional Metadata | |
---|---|
hdl.handle.net/1765/93674 | |
Nederlands Tijdschrift voor Geneeskunde | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Yap, S.-C., Theuns, D., & Szili-Török, T. (2015). Inappropriate ICD shocks after collision with a pole. Nederlands Tijdschrift voor Geneeskunde, 159. Retrieved from http://hdl.handle.net/1765/93674 |