1988
Intracoronary Blood Flow Velocity and Transstenotic Pressure Drop in an Awake Human Being During Coronary Vasodilation
Publication
Publication
Journal of Interventional Cardiology: an international forum for global cardiovascular therapies , Volume 1 - Issue 1 p. 43- 48
The pressure drop over a coronary stenosis and the intracoronary Doppler blood flow velocity were measured at rest and during coronary vasodilation. We report the first observation that confirms the validity of fluid dynamic equations to describe the hemodynamics of a coronary stenosis based on quantitative arteriography in a human being. (J Interven Cardiol 1988:1:1) Copyright
Additional Metadata | |
---|---|
, | |
doi.org/10.1111/j.1540-8183.1988.tb00388.x, hdl.handle.net/1765/94101 | |
Journal of Interventional Cardiology: an international forum for global cardiovascular therapies | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Zijlstra, F., & Serruys, P. (1988). Intracoronary Blood Flow Velocity and Transstenotic Pressure Drop in an Awake Human Being During Coronary Vasodilation. Journal of Interventional Cardiology: an international forum for global cardiovascular therapies, 1(1), 43–48. doi:10.1111/j.1540-8183.1988.tb00388.x |