2003
Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries
Publication
Publication
Circulation (Baltimore) , Volume 107 - Issue 3 p. 381- 383
BACKGROUND: The safety and efficacy of sirolimus-eluting stenting have been demonstrated, but the outcome of patients treated with this novel technology beyond the first year remains unknown. We sought to evaluate the angiographic, intravascular ultrasound (IVUS), and clinical outcomes of patients treated with sirolimus-eluting stents 2 years after implantation. METHODS AND RESULTS: This study included 30 patients treated with sirolimus-eluting Bx Velocity stenting (slow release [SR], n=15, and fast release [FR], n=15) in Sao Paulo, Brazil. Twenty-eight patients underwent 2-year angiographic and IVUS follow-up. No deaths occurred during the study period. In-stent late loss was slightly greater in the FR group (0.28+/-0.4 mm) than in the SR group (-0.09+/-0.23 mm, P=0.007). No patient had in-stent restenosis. At 2-year follow-up, only 1 patient (FR group) had a 52% diameter stenosis within the lesion segment, which required repeat revascularization. The target-vessel revascularization rate for the entire cohort was 10% (3/30) at 2 years. All other patients had < or =35% diameter stenosis. Angiographic lumen loss at the stent edges was also minimal (in-lesion late loss was 0.33+/-0.42 mm [FR] and 0.13+/-0.29 mm [SR]). In-stent neointimal hyperplasia volume, as detected by IVUS, remained minimal after 2 years (FR= 9.90+/-9 mm3 and SR=10.35+/-9.3 mm3). CONCLUSIONS: This study demonstrates the safety and efficacy of sirolimus-eluting Bx Velocity stents 2 years after implantation in humans. In-stent lumen dimensions remained essentially unchanged at 2-year follow-up in the 2 groups, although angiographic lumen loss was slightly higher in the FR group. Restenosis "catch-up" was not found in our patient population.
Additional Metadata | |
---|---|
, , , , , , , , , , | |
hdl.handle.net/1765/10081 | |
Circulation (Baltimore) | |
Organisation | Erasmus MC: University Medical Center Rotterdam |
Sousa, E., Jaeger, J., Popma, J., Costa, M., Sousa, A., Abizaid, A., … Serruys, P. (2003). Two-year angiographic and intravascular ultrasound follow-up after implantation of sirolimus-eluting stents in human coronary arteries. Circulation (Baltimore), 107(3), 381–383. Retrieved from http://hdl.handle.net/1765/10081 |